SmartSite® needle-free system
Topic 1: SmartSite® needle-free valve


Page 2 of 8

Glossary

Resources

Back

Next

Menu

Exit


Using the SmartSite® needle-free valve

1. Prepare Valve

  • Prior to every access, always swab the top of the SmartSite® needle-free valve port with a sterile 70% isopropyl alcohol wipe (1-2 seconds) and allow to dry (approximately 30 seconds). Use of proper aseptic technique helps to reduce the possibility of contamination.
  • For multiple syringes, swab prior to each syringe access.

2. Attach Syringe

  • Insert the syringe tip into the SmartSite® needle-free valve.
  • If syringe has a luer-lock collar, turn syringe clockwise to secure the connection.
  • If syringe has a slip luer, insert and rotate 1/4 turn to secure the connection. Inject or aspirate fluid. Do not leave slip luer syringes unattended.

Important: Do not use needles or blunt cannulas to access the SmartSite® needle-free valve.